07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 9-11
Calibration Slide Clamp Sensor Calibration
3.1 Load a Baxter “s” suffix standard administration set.
3.2 Prime it with fluid.
3.3 Close the pump head door. Ambient and liquid temperatures
must be between 72° and 82° F (22°and 28°C).
3.4 Occlude the tubing upstream of the pump.
3.5 Turn the pump on.
3.6 Set the Primary Rate to 125 mL/hr and Primary VTBI to 50
mL.
3.7 Start the pump and ensure that an upstream occlusion alarm
occurs.
3.8 Press and hold the SILENCE and TOT VOL STATUS keys until
a Failure Identification Code appears in the lower right corner
of the Main Display. The Failure Identification Code should be
4, 5, or 12.
4. Turn off the pump and reassemble the pump door cover.
5. Perform the Operational Checkout procedures in Chapter 10,
“Operational Checkout”.
Slide Clamp Sensor Calibration
Equipment required:
Q DC voltmeter
Q Slide Clamp (part no. 030801011)
Q Screwdriver with flat tip (2.5 to 3 mm width) for potentiometer
calibration. Ceramic or insulated type is preferred.
Q C or E-ring tool
Q Cotton-tipped swabs
1. Turn off the pump.
2. Press the safety clamp arm cover in until the safety clamp latches open.
3. Use the C or E-ring tool to remove the spring retainer from the slide
clamp slot if one is present.
4. Use a cotton-tipped swab dampened with water to clean the safety clamp
sensor. This sensor is located approximately 1/4” in at the upper left of
the slide clamp slot. Repeat with a second dampened swab.