07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 7-1
Chapter 7
Chapter 7 Troubleshooting
Overview
This chapter describes how to find the cause of pump malfunctions. The
section consists of the following:
Q A description of the pump’s alarm and alert messages.
Q A description of the pump’s Failure Identification Codes.
Q A description of the pump’s automatic test modes.
Q A table which lists each failure code, its cause, and ways of correcting
it.
Q A table which describes how to correct problems not represented by
failure codes.
Once the cause of a failure has been determined, perform the corrective action
given in the table. All disassembly/reassembly and calibration procedures for
the pump are in Chapter 8, “Removal/Replacement Procedures” and Chapter
9, “Calibration”. To verify the effectiveness of repairs after they are
completed, perform the Operational Checkout procedures given in Chapter
10, “Operational Checkout”.
! WARNING !
Whenever the pump is in for service the battery’s capacity should be evaluated
because battery performance degrades through its life.
CAUTION
When troubleshooting the pump, do not inject or apply signals unless the
procedures herein specifically instruct you to do so. Damage to the pump or its
subassemblies could result.