07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 1-3
Introduction Safety Summary
To avoid the possibility of electric shock, use caution when the power supply is
connected to the AC input during calibration. Power supply line input is
energized whenever the pump is connected to an AC outlet.
Shock hazard: Do not wear a grounded wrist strap while the pump is open and
connected to AC power.
The pins of the connectors should not be touched. The connectors should not
have objects inserted into them other than the mating connector. ESD
precautionary procedures should be used when inserting the mating
connectors into the connectors.
The power cord must be connected to a 100-120 VAC 50/60 Hz properly
grounded 3-wire receptacle. Do not allow the use of three prong adapters
because this may not provide a reliable earth ground.
Failure to power off the pumps when finished with testing can result in battery
depletion and damage if the pump is operating on battery power, because a
Battery Low alarm will not occur when the pumps are in Configuration/Service
mode.
Follow the cleaning methods described in “Cleaning” on page 6-1 to ensure
proper maintenance of the device.
Do not submerge the pump into liquids of any kind.
Do not clean, disinfect, or sterilize any part of the pump by autoclaving or with
ethylene oxide gas. Doing so may damage the pump and void the warranty. Only
external parts of the pump should be disinfected.
Do not use hard instruments for cleaning. Do not spray cleaners directly into
any openings or gaps in the pump housing.
Do not use the following chemicals on the pump, as they will damage the front
panel: acetoaldehyde, acetone, ammonia, benzene, hydroxytoluene, methylene
chloride, and ozone. Do not use cleaners containing n-alkyl dimethyl ethyl
benzyl ammonium chloride.
Work only at a properly installed electrostatic discharge (ESD) workstation.
Wear a grounding wrist strap, when not connected to AC power, while servicing
the pump.
Do not lay the pump face down on components or tools which could scratch or
damage the keypad. Do not lay the pump face down with door assembly open.
Ensure that wires are not pinched or overstressed.
When troubleshooting the pump, do not inject or apply signals unless the
procedures herein specifically instruct you to do so. Damage to the pump or its
subassemblies could result.
Tighten fasteners to the torque values specified in the procedures. Do not
overtighten any fastener.
Ensure that connectors are mated and observe the correct orientation.