7 Alerts
7-6 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-729
CHECK VTBI Yellow No flow PRI or SEC
START
Enter a VTBI within the range selected through
the configuration option. The pump cannot be
started when Hi is displayed in a VTBI display.
INSERT
SLIDECLAMP
Yellow No flow or
flow if started
Closed door,
PRI or SEC
START
Slide clamp is not loaded into the slide clamp
slot although the set is loaded into the pump.
COM TIME OUT Yellow No change or
no flow
None Communication timeout period has elapsed. No
communication has occurred between the pump
and the host computer during the power-up
default time period (60 seconds) or during the
time period most recently specified by the host
computer (1-300 seconds). The computer and the
pump must maintain periodic and successful
communication in order to avoid this timeout
alert. See the Programmer’s Manual for further
information.
EXT COMM
ERROR
Yellow No change None The controlling or monitoring computer is
sending multiple queries or commands to the
pump without waiting for the pump’s replies.
The condition has been been caused by the
computer or the pump CPU board. To clear the
alert, press OPTIONS to return the pump to
computer control (if appropriate), or press
SILENCE or PUMP 1 or PUMP 2 to control
the pump(s) manually. Notify the technical
personnel responsible for the computer. When
the alert accurs, disconnect the cable from the
communications port. Reprogram the pump(s)
for manual operation.
PGM DELIV
ENTER PGM
Yellow No flo w PRI or SEC
START
An attempt was made to start the PDP before a
program was entered. Enter a program or press
OPTIONS to leave the PDP mode.
PGM DELIV
REVIEW PGM
Yellow No flo w PRI or SEC
START
An attempt was made to start a PDP profile prior
to reviewing all the programmed steps. Press
NEXT until the first step is again displayed or
press OPTIONS to leave the PDP mode.
PGM DELIV
CLEAR ALL?
Yellow No flo w CLR The CLR key was pressed after entering the PDP
mode to clear all steps. Do one of the following:
press CLR to erase the profile; press NEXT to
review the rest of the profile; press PRI RATE
or PRI VTBI to modify this step; press
OPTIONS to leave the PDP mode.
Table 7-2 Alert Messages — continued
Alert Message LED Flow Status
Key
Pressed Alert Condition