7 Automatic Test Modes
7-10 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-729
Automatic Test Mode 7: Pumping Sensor Monitoring Mode
In this test mode, the motor current and occlusion sensors may be monitored
during pumping. During pumping, the motor current is displayed in the PRI
VTBI location, the upstream occlusion detector reading in the SEC RATE
location, and the downstream occlusion sensor reading in the SEC VTBI
location. When the pump is stopped, all of the displayed values return to
normal.
Automatic Test Mode 8: Air Sensor Test Mode
Upon entering this mode, the PRI RATE is set to 30, PRI VTBI = 100, SEC
RATE = 600, and SEC VTBI = 100. Pressing the
TIME key selects the pump’s
air sensor sensitivity (MIN or NORM) without using the configuration option.
The air sensor selection is displayed in the message display when the pump is
started. All other keys function normally except for the
TIME key. Rate and
VTBI information for both primary and secondary mode may be changed.
Automatic Test Mode 9: Elapsed Time Test Mode
The elapsed pumping time is displayed in the first line of the message display
in this test mode. Momentarily pressing the
TIME key selects two
pre-programmed primary mode test conditions. The first is a PRI RATE of
125 mL/hr with a VTBI of 100 mL. The second is a PRI RATE of 499 mL/hr
with a VTBI of 100 mL. The elapsed time counts only when the pump is
running. Stopping the pump and then restarting it resets the timer to zero. All
other keys function normally except for the
TIME key. When the VTBI has
been delivered, the pump enters KVO mode.
Automatic Test Mode 0: Downstream Occlusion Test Mode
In this mode, pressing the TIME key allows you to select the pump’s
downstream occlusion level sensitivity without entering the configuration
mode. The selected occlusion sensor level (1, 2, or 3) is momentarily shown
in the second line of the message display until the test mode is exited. After
exiting test mode, the downstream occlusion sensor level reverts to the
original configuration setting. Rate and VTBI information may be
programmed manually.
! WARNING !
This is an automatic test mode for testing the pump during manufacturing only.
This mode must not be used on patients.