07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 7-5
Troubleshooting Alerts
Alerts
Alert messages call attention to a condition that require operator
intervention in the near future, or indicate that a procedure has been initiated
which requires that the operator complete a sequence of keystrokes. These
alerts are generally cleared by the operator taking the appropriate action.
Table 7-2 lists the various alerts and possible causes.
Table 7-2 Alert Messages
Alert Message LED Flow Status
Key
Pressed Alert Condition
STOPPED Yellow No flow None The pump has been left in the STOPPED mode
for more than two minutes.
KVO PRI VTBI = 0 Green,
Yellow
KVO rate None Primary VTBI has been delivered and the pump
has switched to 5 mL/hr KVO rate (or
programmed rate, whichever is lower).
NEW RATE Green,
Yellow
No change
until
procedure is
completed
PRI or SEC
RATE
Primary or Secondary flow rate is being changed
while pump is running. Pump does not exit this
alert condition until the appropriate START key
is pressed.
PRI RATE = 0 Yellow No flow PRI or SEC
START
The pump cannot be started without entering a
primary flow rate.
BATTERY LOW
with audible alert
Green,
Yellow
No change None Battery needs recharging. The pump stops
operating in approximately fifteen minutes or
longer unless it is plugged into an AC source.
SEC PROGRAM Green,
Yellow
No change SEC RATE
or SEC
VTBI
Secondary (piggyback) information is being
programmed into the pump while it is running in
primary mode. Pump does not exit this alert
condition until SEC START is pressed.
SEC RATE = 0 Yellow No flow SEC
START
A secondary (piggyback) infusion cannot be
started unless a secondary flow rate is input.
SEC VTBI = 0 Yellow No flow SEC
START
A secondary (piggyback) infusion cannot be
started unless a secondary volume to be infused
has been input.
SEC COMPLETE Yellow No change None The pump has completed infusing the secondary
VTBI and has switched over to the primary
infusion settings. Pump does not exit this alert
condition until any enabled key is pressed.
FLOW RATE Yellow No flow PRI or SEC
START
Enter a rate within the range selected through the
configuration option. The pump cannot be
started when Hi or Lo is displayed in a rate
display.