1 Safety Summary
1-2 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-729
The primary rate of infusion is selectable from 1 to 1999 mL in 1 mL/hr
increments and 1 to 99.9 mL/hr in 0.1 mL/hr increments. The secondary rate
is selectable from 1 to 999 mL in 1 mL/hr increments and 1 to 99.9 mL/hr in
0.1 mL/hr increments. The volume to be infused (VTBI) is also selectable
from 1 to 9999 mL in 1 mL increments and 1 to 99.9 mL in 0.1 mL
increments.
The total volumes infused from primary and secondary programs are added
together and accumulated and can be displayed on demand. The primary and
secondary VTBIs are independently decremented and displayed. Upon
completion of the primary VTBI, the pump automatically switches to a keep
vein open (KVO) rate. When the pump is started on a secondary rate and
VTBI, the pump changes to the primary rate when the secondary program is
completed. Either pump may be stopped at any time by pressing the
STOP
key unless the device is in the lockout mode.
Safety Summary
General precautions to observe while servicing the pump are shown below.
Warnings and cautions related to personal safety are summarized in this
section.
! WARNING !
This device should be repaired only by Baxter authorized service personnel or
Baxter-trained hospital biomedical engineering personnel.
! WARNING !
If the pump has been dropped or appears to be damaged, it should be taken out
of service and carefully inspected.
! WARNING !
Use only Baxter replacement/service parts. Installation of parts other than those
available from Baxter may affect product performance. The use, by any party, of
non-Baxter parts in the repair, renovation, repackaging or restoration of a
Baxter device means that person or entity is the manufacturer of that device.
Any liability arising from the use of non-Baxter parts is assumed by the entity
performing such work. Baxter’s warranty is immediately null and void where
non-Baxter parts are used to repair Baxter devices.
! WARNING !
Batteries are intended to be used as a back-up power source ONLY when AC
power is not available. The pump should be plugged in to AC power to maintain
battery charge during use and storage whenever possible.
! WARNING !
If the charging state of the batteries is unknown or the pump has not been
plugged in to AC power for at least 16 uninterrupted hours then the pump must
be charged for at least 16 uninterrupted hours of charging prior to releasing it
for use.
! WARNING !
Whenever the pump is in for service the battery’s capacity should be evaluated
because battery performance degrades through its life.
! WARNING !
Ensure that the pump is powered off and unplugged from the AC receptacle
before performing maintenance procedures unless directed otherwise by the
procedure in this manual.