A-2 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-729
Pump Serial Number: _____________________________
Printed Name Signature Date
Tester: _____________________________ _____________________________ ____________
Reviewer (optional): _____________________________ _____________________________ ____________
N/A _________
Calibrated Syringe Due Date/ID: ________________________________
Door Open and Alarm Volume Test
Pump 1 "DOOR OPEN" message displays
Alarm LED flashes
Audible alarm sounds
No R Yes R
Pump 2 "DOOR OPEN" message displays
Alarm LED flashes
Audible alarm sounds
No R Yes R
Slide Clamp Mechanism Test
Pump 1
"INSERT SLIDECLAMP" message displays No R Yes R
Pump 2
"INSERT SLIDECLAMP" message displays No R Yes R
Air Alarm Test
Pump 1
NORM
(50μL) No air detected Fail R Pass R
(110μL) Air detected Fail R Pass R
MIN
(25μL) No air detected Fail R Pass R
(85μL) Air detected Fail R Pass R
Pump 2
NORM
(50μL) No air detected Fail R Pass R
(110μL) Air detected Fail R Pass R
MIN
(25μL) No air detected Fail R Pass R
(85μL) Air detected Fail R Pass R