5 Modifying the Configuration Option Settings
5-4 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-729
11. Flow Check: Determines whether the flow check display appears
during device operation, or only when the decimal point key and TOT
VOL/STATUS keys are pressed simultaneously.
1: OFF
2: ON
OFF
12. Baud Rate: Determines the baud rate for normal communications
between the pump and a computer. The baud rate is 9600 when the
pump is in modify configuration mode.
1: 300
2: 1200
3: 2400
4: 4800
5: 9600
9600
13. Computer Control: Determines the type of computer control option
available at power up.
1: Disabled
2: Off with Alarm. The pump
drops out of remote control
when an alarm occurs.
3: On with Alarm. The pump
remains in remote control
when an alarm occurs. See
the Programmer’s Manual
2M8963C for additional
information.
Disabled
14. Hospital Area Designator: Determines the hospital area designator
to be displayed upon power ON for 3 seconds. These messages can be
redefined using the computer communications feature.
0: no message
1: NICU
2: PICU
3: MED/SURGICAL
4: TRAUMA/BURN UNIT
5: OPER ROOM
6: CARDIAC/ICU
7: SURGICAL/ICU
8: ICU
9: ONCOLOGY
0
15. Close Clamp: Determines whether or not the CLOSE CLAMP
message appears with the DOOR OPEN message.
1: OFF
2: ON
ON
16. Insert Clamp: Determines whether or not the slide clamp loading
feature is enabled. The spring retainer must be removed when this
feature is enabled. See “Spring Retainer Removal” on page 8-22 and
“Spring Retainer Installation” on page 8-22.
Note: The alert or alarm option is available only on pumps running
software versions 1.09 or later.
1: OFF (The spring retainer
should be installed.)
2: ON for software versions
earlier than 1.09 or ALERT
for software versions 1.09 or
later (An audible alert occurs
and the message INSERT
SLIDECLAMP is displayed.
The pump continues
pumping.
3: ALARM (An audible
alarm occurs and the
message INSERT
SLIDECLAMP is displayed.
The pump does not operate.)
ALARM
Table 5-1 Configurable Settings — continued