Chapter 3 Parameters
Mini-Switch LB 471
49
Save& Load / Reset
Save/Load parameter set or initiate reset
(Only in the professional mode)
Input
Function
0
Save parameter set
1
Load parameter set
99
Initiate software reset
999
Initiate software reset and reset
parameters to factory setting
Save parameter set
The present parameter set is saved to a separate memory on the
EVU. We recommend storing the parameter set if you want to save
the settings you have defined in order to access them later. The
saved setting can be invoked again after a reset.
Load parameter set
A parameter set previously saved on the EVU can be loaded again;
the current parameters will be overwritten.
Restart
Software restart without changing the instrument parameters. If
mysterious malfunctions occur, the CPU can be reset in this code.
Reset
Same as above but with additional parameter reset to factory set-
ting. This function is comparable with turning on the power supply
with the Clear button pushed down (see page 72).
C o d e 7 2