EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB4710-080 - Page 213

Default Icon
227 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6 Error Messages
Mini-Switch LB 471
Error
number
Error
sub-code
Designation
Display
Reaction of
measurement
Alarming relay
Cause
Remedy
01
XX
Hardware error
Err
Halt
Error relay
Fault in a hardware compo-
nent or the board
Replace evaluation unit
04
01
Watchdog reset
!!
RUN
Dependent on
code 44
High electrical interference
or faulty hardware
Check power line for possible interfer-
ence
If necessary, replace evaluation unit
02
Watchdog reset
>2x in 10s
Err
Halt
Error relay
As above, but the malfunc-
tion occurred several times
within 10s.
Check power line for possible interfer-
ence
If necessary, replace evaluation unit
05
01
No count rate
Err
Halt
Error relay
No counting pulse occurred
within 10s (FSK) or 120s
(GMZ)
Replace probe
06
01
HV < -20%
Err
Halt
Error relay
HV is 20% below the HV
start value of code 54
Check HV and standard HV.
If necessary, update standard HV.
If necessary, replace detector.
02
HV > +40%
Err
Halt
Error relay
HV is 40% above the HV
start value of code 54
Check HV and standard HV.
If necessary, update standard HV.
If necessary, replace detector.
03
HV < 500V
Err
Halt
Error relay
HV is at lower limit at 500V
Check HV.
If necessary, replace detector.
04
HV > 1300V
Err
Halt
Error relay
HV is at upper limit at 1300V
Check HV.
If necessary, replace detector.
09
01
Detector tempera-
ture exceeded
(only FSK)
!!
RUN
Dependent on
code 45
Detector heating up > value
in code 46
(only FSK)
Reduce ambient temperature on probe:
e.g. through heat deflection plate or
water cooling.
10
01
EVA temperature
exceeded
!!
RUN
Dependent on
code 47
EVU heating up > value in
code 48
Reduce ambient temperature on evalu-
ation unit
11
01
Communication
error
Err
Halt
Error relay
Faulty communication de-
tected (only FSK)
Replace probe or evaluation unit or
eliminate line interruption
17
01
Date error
!!
RUN
Dependent on
code 44
Date error detected
Check date in code 01/02, if necessary
update it
22
01
Interference radia-
tion detection
!!
Waiting time
code 41
Dependent on
code 42
Interference radiation or
faulty calibration. Corre-
sponds to error messages
35.04 to 35.08, will be
checked during operation.
See code 40 on page 38.

Table of Contents