23
RECALL DEFAULT:
Restores the manufacturer
default
settings
KEY LOCK:
While exiting the setting screen after enabling
KEY
LOCK, all keys except the key
are locked.
Only this key can disable KEY LOCK.
*Simultaneously pressing both and keys in the main
screen can also lock keys.
*While entering KEY LOCK state, the display will show an LCK
indicator in the bottom right corner.
POWER ON STATE:
Users can set the output state
of the supply
when it is powered on. Wh
en OFF is
selected,
the power on state configuration
will not be recalled. If LAST is selected,
then
at power on the supply will configure to the
last settings before it was turned off
previously. If USER (user defined)
is
selected, a prompt will ask for setting
output
voltage, current, and output state.
Once set,
these values will then apply during
the next
power up.
IP CONFIG:
STATIC : User can input IP address
DHCP : not supported
IP ADDRESS:
Users can enter a static IP address here.
=
.
=
FF
F
=
,
=
.
F
=
,
FF
L
=
5
F
52=
= F
552. 552. 552.
F
L