64
:CLEar clear program n parameters
:ALL clear all program parameters
:NEXT <NR1 | ? > ser or return next program number (1 -
10 , 0 for end)
:REPeat <NR1 | ? > set or return repeat times (max.
50000)
:RUN<Boolean | ? > set or query program on/off state
:SAV save program parameters
:STEP < Boolean | ? > set or return step number
:CURRent <NRf | ? > set or return step n current setting
:ONTimet <NRf | ? >
set or return step n output time (0.050 -
20000S)
:VOLTage <NRf | ? > set or return step n voltage setting
:TOTA[l] <NR1 | ? > set or return program n total step
numbers (max. 150)
PROTection Protection Subsystem
? return protect state
:CCCV <Boolean | ? > set or return CC to CV protect state
:CLEar Resets latched protection
:CVCC <Boolean | ? > set or return CV to CC protect state
:OCP <Boolean | ? > set or return over-current protect state
:LEVel <NRf | ? > set or return over-current protect value
:OPP <Boolean | ? > set or return over-power protect state
:LEVel <NRf | ? > set or return over-power protect value
:OVP <Boolean | ? > set or return over-voltage protect state
:LEVel <NRF | ? > set or return over-voltage protect value
PS Parallel/Series Subsystem