95
THEN/ELSE Statement Description
Starts the next phase. Starts the next phase.
Go to Phase 1 Starts Phase 1.
Go to Phase 2 Starts Phase 2.
Go to Phase 3 Starts Phase 3.
Go to Phase 4 Starts Phase 4.
Go to Phase 5 Starts Phase 5.
Go to Done Starts Phase Done.
REPEAT 0 times Repeats actions in the current phase the number of
times specified.
Available settings:
1 to 8
FOREVER
. The program continues forever. See the “IF
Loop Count<Max” and “IF Loop Count>Max”
statements, above.
REST for 000:00:00 Applies no current for the specified time period (in
hhh:mm:ss format). The
RESTING
message (code 19) is
displayed with time counting down.
After resting, the program goes to the next phase.
FAIL Fails the battery. The battery is electrically disconnected,
the
FAIL
light comes on, and the
WARNING
message (code
16) appears.
Phase Done
Perform one of three different actions in the final phase of the custom program.
Action Description
ReadyNo Charge Ends the program. The green
READY
light is on, and no
trickle charge is applied.
ReadyTrickle Charge Trickle-charge the battery using the trickle-charge rate in
the extended C-Code. Includes the standby maintenance
cycle, which performs a discharge/ charge every 30 days
(180 days for SLA).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cadex Electronic C7000-C Series and is the answer not in the manual?

Cadex Electronic C7000-C Series Specifications

General IconGeneral
BrandCadex Electronic
ModelC7000-C Series
CategoryMeasuring Instruments
LanguageEnglish