4. In the dialog box that appears, enter these settings:
Connect using. From the list, select Direct to Comx, where x is the number of the serial port to
which the analyzer is connected.
•
Bits per second: 9600
•
Data Bits: 8
•
Parity: None
• Stop Bits: 1
•
Flow Control: None
5. Click OK.
The HyperTerminal screen appears. The bottom left-hand corner of the screen should
display the message “Connected h:mm:ss”.
6.
Check that HyperTerminal is communicating with the Cadex analyzer by
inserting an adapter in any station on the analyzer. A text string similar to
this one should appear on-screen:
0,2," ","01/25/2001","095126",201,"0\80"
If nothing appears on-screen, check the HyperTerminal properties:
Click File, and then click Properties.
Verify that the serial communication port number is correct.
Click Configure, and verify that the serial port settings match the Cadex analyzer serial port
settings.
If the Properties settings are correct, check the serial cable and connections, and verify
the serial port settings on the computer and battery analyzer.
Using the Data Log
Once the connection file has been set up, use that file to open HyperTerminal and log
Cadex analyzer data. Save Cadex analyzer data logs by capturing them as text files, which