0 = Auto 8 = RunTime
1 = Charge 9 = Self-Discharge
2 = Prime 10 = LifeCycle
3 = Custom1 11 = Discharge Only
4 = Custom2 12 = Extended Prime
5 = Custom3 13 = Boost
6 = Custom4 14 = QuickTest
7 = OhmTest 15 = Q-Learn
16 = Learn
Battery data. In certain cases, event strings also contain a battery data field. For
example, when an automatic OhmTest is run as part of a program, the event string
generated includes the OhmTest results in m?.
Sample text strings
Here are some sample data logging text strings followed by explanations.
0,2," ","01/24/2001","085120",201,"0\80"
a) A Battery Adapter was inserted (Code 201) in station 2 on January 24,
2001 at 8:51 a.m. The current C-Code specifies the Auto program with
a target capacity of 80% (0\80 in the final field). Battery ID is not yet
available.
0,2," ","01/24/2001","085121",20,"0\80"
b) A battery was inserted in the adapter (Code 20).
0,2,"CDX01 ","01/24/2001","085140",11,"0\80"
c) Processing of the current program began (Code 11). The battery ID is
CDX01.
0,2,"CDX01 ","01/24/2001","085140",250,"2\1416\398\21",""
d) Normal processing (Code 250) of battery CDX01. The battery is
charging (process = 2), terminal voltage is 1416mV, charge current is