Cadex Electronics Inc. C7000-C Series Battery Analyzers .
To turn Automatic OhmTest on or off
1.
From the global display, press
MENU
.
2.
Press 5 or 6 to scroll to
PROGRAM SETTINGS
and then press
ENTER
.
3.
Press 5 or 6 to scroll to
AUTOMATIC
O
HM
T
EST
and then press
ENTER
.
4.
Press 5 or 6 to scroll to the required setting:
• YES
(default) sets the OhmTest to run automatically before every charge
cycle (except during custom programs).
• NO turns off the automatic OhmTest.
5.
Press
ENTER
.
6.
Press 5 or 6 to scroll to
YES
to accept changes and press
ENTER
.
7. Press
ESC
twice to return to the global display.
Optimizing Charge Time or Capacity
The end of charge conditions can be modified, (see Appendix B – Charge Algorithms,
page 135).
To select the charge optimization option
1.
From the global display, press
MENU
.
2.
Press 5 or 6 to scroll to
PROGRAM SETTINGS
and then press
ENTER
.
3.
Press 5 or 6 to scroll to
CHARGE OPTIMIZATION
and then press
ENTER
.
4.
Press 4 or 3to move to between N
I
C
D AND
NIMH
OPTIMIZATION
and SLA
AND
LI O
PTIMIZATION
:
5. Press 5 or 6 to the required setting (see Appendix B – Change
Algorithms, page 135)
• TIME
(default).
• CAPACITY
.
6.
Press
ENTER
.