39
Sequence: The battery is tested to ensure that it has 20% - 90% State of Charge. If not, it
applies a charge or discharge for five minutes. The program then performs a sequence of
resistance tests, partial charges, discharges and rest. Test results are compared to the
matrix to determine the battery state of health.
Results: The Global and detailed display shows the battery condition as “Excellent”,
“Good”, “Marginal” or “Poor”. Detailed display also shows average battery or cell voltage,
analyzer charge or discharge current in mA, OhmTest results, battery temperature (if
applicable) and duration of service.
Duration: Approximately 5 minutes if the battery SoC (state of charge) is between 20%
and 90%. Batteries with a lower or higher SoC require additional time to bring their SoC to
between 20% and 90%. For most batteries, this takes about 5 minutes. However, with
high-capacity batteries, it may take longer.
Custom 1, 2, 3, and 4
By default, custom programs are blank when the Cadex analyzer is shipped from the
factory (see Chapter 10 - Custom Programs, page 91).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cadex Electronic C7000-C Series and is the answer not in the manual?

Cadex Electronic C7000-C Series Specifications

General IconGeneral
BrandCadex Electronic
ModelC7000-C Series
CategoryMeasuring Instruments
LanguageEnglish