Chapter 12
System Calibration
There are two kinds of calibration that should be performed on the Cadex analyzer on a
regular basis: voltage calibration and adapter (or station) calibration.
Calibrating Voltage
Voltage calibration should be performed yearly and a calibration kit is required.
The C7000-C Series of Battery Analyzers uses Calibration Kit P/N 92-770-0212 (See
Appendix D - Parts and Accessories, page 141). This kit CANNOT be used on the
non-C series analyzers: C7000, C7200, C7400, C7000ER and C7400ER.
To calibrate the voltage
1.
Remove all Battery Adapters.
2.
From the global display, press
MENU
.
3. Press
5
or
6
to scroll to SYSTEM SETUP and then press ENTER.
4. Press
5
or
6
to scroll to CALIBRATE VOLTAGE and then press ENTER .
5.
Firmly insert the four voltage calibration adapters, one into each station. (If
calibrating a C7200, only two voltage calibration adapters are necessary)
6.
Wait for the prompt, and then remove the calibration adapters.
7.
Press
MENU
to return to the global display.
8.
Press
FN
, and then press 7. The “#” sign appears in the upper right corner
of the startup screen, indicating successful completion of voltage calibration.
Servicing the Voltage Calibration Adapters
Return the voltage calibration adapters to the factory every three years for authorized
calibration. Contact Cadex for a Return Authorization number.