1. Press
5
to move to the next phase.
2.
Repeat steps 6 – 1 for phases 2 – 5.
Once Phases 1 through 5 have been programmed, the cursor appears on Phase Done.
3. Program the action for Phase Done:
Press
4
to move to Cycle 1.
Press 5 or 6 to select the action to be performed. (There is only one cycle for Phase
Done.)
1.
After all the settings have been entered, press
ENTER
.
2.
Enter a name for the program, if desired:
Press 5 or 6 to scroll through upper case letters, lower case letters, punctuation
marks, numbers, and a space (which looks like an underscore) to select characters for the
program name.
Press 4 or 3 to move the cursor between character positions within the program name.
There are seven positions available.
3.
Press
ENTER
.
4.
Press 5 or 6 to scroll to
YES
to accept changes and press
ENTER
.
5.
Press
ESC
twice to return to the global display.
Sample Custom Program
NiCd or NiMH batteries that do not meet the target capacity should always be
reconditioned. If the battery does not meet the target capacity after reconditioning, the
program fails the battery.
Phase Action
Phase 1 Charge the battery.
Phase 2 Discharge to find the battery capacity. If the capacity meets the target
capacity, complete the program (Phase Done). If the target capacity is not
met, apply the recondition cycle (Phase 3).
Phase 3 Recondition and charge the battery.