Cadex Electronics Inc. C7000-C Series Battery Analyzers .
Results: Global display shows final battery capacity percentage. Detailed display shows
battery capacity percentages for the last three cycles performed as well as average
battery or cell voltage, analyzer charge or discharge current in mA, OhmTest results,
battery temperature (if applicable) and duration of service.
Batteries in good condition should be greater than 80% or the target capacity. If batteries
fail, check the fault codes for details. If the capacity is low, the battery should be Primed
again. In some cases, two or three Prime programs may be required to fully form a
battery.
Duration: Approximately 5 to 10 hours for NiCd and NiMH, from 40 to 80 hours for SLA,
from 12 to 25 hours for Li-ion using default C-Codes.
QuickSort
Notes: QuickSort is available only on 16k Battery Adapters. These adapters are marked
with a ‘16K’ label on the rear of the adapter. The QuickSort program can be selected only
when the Li-ion C-Code is selected.
The battery must have at least 40% State of Charge. Charge a battery fully if you are
unsure of the battery’s State of Charge.
Function: This program works for single cell Li-ion, 3.6V, and 500mA to 1500mAh
batteries and with the FlexArm adapter (07-110-0180).
Sequence: QuickSort consists of several test pulses followed by an OhmTest. An
approximation to the battery’s condition is then calculated. There is no maintenance
charge after this program is completed.
Results: The Global and detailed display shows the battery condition as “Good”, “Low” or
“Poor”. Detailed display also shows average battery or cell voltage, analyzer charge or
discharge current in mA, OhmTest results, battery temperature and duration of service.
Duration: 30 seconds.