EasyManua.ls Logo

Cadex Electronic C7000-C Series - Performing an Automatic Ohmtest

Cadex Electronic C7000-C Series
163 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
The Cadex analyzer starts the program using the active C-Code without asking to enter the
Battery ID or verify the C-Code.
To verify C-Code or enter Battery ID before battery service begins
1.
From the global display, press
MENU
.
2.
Press 5 or 6 to scroll to
OPTION CONTROLS
and then press
ENTER
.
3.
Press 4 or 3.to scroll to
ASK FOR BATTERY ID
#
4.
Press 5 or 6 to scroll to the required setting:
YES
sets the Cadex analyzer to ask for the Battery ID number.
NO
(default) sets the Cadex analyzer not to ask for Battery ID number.
5. Press 4 or 3.to scroll to
ASK FOR BATTERY C
-
CODE
5.
Press 5 or 6 to scroll to the required setting:
YES
sets the Cadex analyzer to ask the user to verify C-Code settings.
NO
(default) sets the Cadex analyzer not to ask for the battery
C-Code.
6.
Press
ENTER
.
7.
Press 5 or 6 to scroll to accept changes and press
ENTER
.
8.
Press
ESC
twice to return to the global display.
Performing an Automatic OhmTest
Every time the Cadex analyzer runs a charge cycle, it can automatically perform an
OhmTest on the battery as well (except when the charge cycle is being performed as part
of a custom program). This test can be turned off.
Note Automatic OhmTest cannot be turned on or off using the following programs:
QuickTest, Q-Learn, Learn and Life-Cycle. OhmTest is included in these programs
and the settings cannot be changed.

Table of Contents