The Cadex analyzer starts the program using the active C-Code without asking to enter the
Battery ID or verify the C-Code.
To verify C-Code or enter Battery ID before battery service begins
1.
From the global display, press
MENU
.
2.
Press 5 or 6 to scroll to
OPTION CONTROLS
and then press
ENTER
.
3.
Press 4 or 3.to scroll to
ASK FOR BATTERY ID
#
4.
Press 5 or 6 to scroll to the required setting:
• YES
sets the Cadex analyzer to ask for the Battery ID number.
• NO
(default) sets the Cadex analyzer not to ask for Battery ID number.
5. Press 4 or 3.to scroll to
ASK FOR BATTERY C
-
CODE
5.
Press 5 or 6 to scroll to the required setting:
• YES
sets the Cadex analyzer to ask the user to verify C-Code settings.
• NO
(default) sets the Cadex analyzer not to ask for the battery
C-Code.
6.
Press
ENTER
.
7.
Press 5 or 6 to scroll to accept changes and press
ENTER
.
8.
Press
ESC
twice to return to the global display.
Performing an Automatic OhmTest
Every time the Cadex analyzer runs a charge cycle, it can automatically perform an
OhmTest on the battery as well (except when the charge cycle is being performed as part
of a custom program). This test can be turned off.
Note Automatic OhmTest cannot be turned on or off using the following programs:
QuickTest, Q-Learn, Learn and Life-Cycle. OhmTest is included in these programs
and the settings cannot be changed.