Cadex Electronics Inc. C7000-C Series Battery Analyzers .
Lights (LEDs)
Light
Status Explanation
RUN (yellow)
On
Service in progress.
READY (green) On
Service has completed; Battery has passed the
service.
Flashing
Cold battery. Service resumes when battery warms
up (code 12).
FAIL (red) On
Battery failed.
Flashing
Hot battery. Service resumes when battery cools
(code 13).
All
Flashing
randomly
System failed. Turn the analyzer off and then on.
Contact Cadex if the condition persists.