Cadex Electronic C7000-C Series - Termination Condition on Dt;Dt; Saving Program and Target Capacity Settings

Cadex Electronic C7000-C Series
163 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71
7. Press
5
or
6
to scroll to YES to accept changes and press ENTER.
8.
Press
ESC
twice to return to the global display.
Termination Condition on dT/dt
The Termination Condition on DT/DT parameter sets the temperature a battery can reach
before it will terminate or reduce the charge current. If a battery temperature rises more
than the
D
T/
DT
the charge current is either reduced or charge is completed (see Appendix
B Charge Algorithms, page 135).
To select the charge optimization option
1.
From the global display, press
MENU
.
2. Press 5 or 6 to scroll to
PROGRAM SETTINGS
and then press
ENTER
.
3.
Press 5 or 6 to scroll to
D
T/
DT TERMINATION
and then press
ENTER
.
4.
Press 5 or 6 to the required setting (see Appendix B Charge
Algorithms, page 135)
2
O
C
RISE PER
: 2
MINUTES
2
O
C
RISE PER
: 3
MINUTES
(default)
2
O
C
RISE PER
: 4
MINUTES
5.
Press
ENTER
.
6.
Press 5 or 6 to scroll to
YES
to accept changes and press
ENTER
.
7. Press
ESC
twice to return to the global display.
Saving Program and Target Capacity Settings
By default, when program and target capacity settings are changes, they are saved in the
C-Code and stored in the Battery Adapter.
The Cadex analyzer can save these changes to the battery station rather than to the C-
Code. Settings saved to the station are independent of setting Battery Adapter inserted in
the station. Program and target capacity settings programmed into the analyzer station
will override the C-Codes programmed in the adapter.