EasyManuals Logo

Cadex Electronic C7000-C Series User Manual

Cadex Electronic C7000-C Series
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
Cadex Electronics Inc. C7000-C Series Battery Analyzers .
118
By Detailed Message
Detailed
Message
Code Reasons Solutions
ADAPTER DATA
INVALID
209
Invalid value in selected C-
Code.
The C-Code is not compatible
with the firmware.
There may be a defect with the
adapter memory
Delete and re-enter the C-Code
reset the extended C-Code to
default values.
Contact Cadex to upgrade
firmware.
ADAPTER INSERTED
201
Adapter is detected on a
station. This message appears
momentarily.
If the message persists, Press
Esc.
ADAPTER NOT SET
UP
208
Adapter is not programmed
with a C-Code, or the C-Code is
invalid.
Delete and re-enter the C-Code.
Contact Cadex to upgrade
firmware.
ADAPTER REMOVED
202
Adapter is removed from a
station.
If message persists, press
ESC
.
ADAPTER SETUP
UPDATED
206
Cadex BatteryShop has
updated c-code in the Battery
Adapter.
If message persists, press
ESC
.
BAD ADAPTER
210
The analyzer does not
recognize adapter.
Remove and re-insert the
adapter.
Delete and reenter the C-Code.
Clean the adapter contacts.
Contact Cadex to upgrade
firmware.
If message persists, replace
adapter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cadex Electronic C7000-C Series and is the answer not in the manual?

Cadex Electronic C7000-C Series Specifications

General IconGeneral
BrandCadex Electronic
ModelC7000-C Series
CategoryMeasuring Instruments
LanguageEnglish