Cadex Electronics Inc. C7000-C Series Battery Analyzers .
84
Process or status. This number represents the process or status of the current program
as listed in Appendix A, Messages and Warnings (see By Code).
Voltage. This is the terminal voltage of the battery in mV.
Current. This is the charge or discharge current being applied to the battery in mA. If the
current number is positive, the battery is being charged; if it is negative, the battery is
being discharged.
Battery temperature. This is battery temperature in degrees Celsius. If temperature
sensing is disabled in the C-Code, this value is always 0.
Back-slash characters separate the four elements.
Battery capacity. In most cases, this field displays the current (or most recent) battery
capacity and the previous battery capacity, both as percentages of the nominal
battery capacity in the C-Code. A back-slash character separates the two capacities.
Current battery capacity is updated (increased) during discharge cycles.
Note If running QuickTest, this field displays the battery SoH (State of Health) as a
percentage, followed by the ASCII value of the confidence grade for the matrix (A = 65, B
= 66, C = 67).
The bottom text string in Figure 9 is a processing string. It indicates that the battery in
station 1 (ID CDX0095468) is processing normally (event 250), charging (process 2) with
a terminal voltage of 1345mV, a charge current of 801mA, and a battery temperature of
28°C. The most recent battery capacity achieved was 74%; the previous capacity was
68%.
Event strings contain one or two fields in addition to the first six:
Program and target capacity. This field displays a one- or two-digit number
representing the current program and the target capacity from the C-Code. A back-
slash character separates the two elements.
Program codes are:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cadex Electronic C7000-C Series and is the answer not in the manual?

Cadex Electronic C7000-C Series Specifications

General IconGeneral
BrandCadex Electronic
ModelC7000-C Series
CategoryMeasuring Instruments
LanguageEnglish