Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
Figure 3-20 Parallel 3-Phase Delta Connection
Users can set POSITION to determine master and slave as setting 3-phase parallel mode,
POSITION setting is 1 ~ 5. If POSITION = 1, it means the AC load will be set as master. If
POSITION = 2 ~ 5, it means AC load will be set as slave. In addition, master needs to set
TOTAL UNIT quantity to ensure loading power.
It is needed to start setting from slave of A phase and master/slave of B/C phase as setting
3-phase parallel mode. That is to say, users need to set AC load to slave and Master of B/C
phase setting is completed, set master of A phase at last thus master of A phase can confirm