Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
12. Press 2 , 0 then press ENTER to change the value to “00:20:00”.
13. Move the cursor by pressing ( ◄, ► ) to the command line of “CUT OFF V=”.
14. Press 1 , 1 , 0 then press ENTER to change the value to “110.00”.
15. After setting the parameter of the load, the LOAD ON/OFF key should be pressed to turn
on the load, so that the load can sink the current. (For detailed information please refer
section
3.4.5.)
If there is error during the procedure of inputting value, press CLEAR to clear
the value at any time and then input again.
For switching to CF only mode, the procedures are as follows.
1. Press CONF./LOCAL key.
2. Press 1 , then press ENTER to enter SETUP.
3. Move the cursor ( ▲, ▼ ) to “1-5. CF/PF SETTING”.
4. Press the direction key ( ◄, ► ) or rotate the rotary knob to choose “CF”, then press
ENTER.
5. Then press CC to back to main page.
Users can press the arrow key to select the setting parameters on the LCD display, including
current RMS value, CF, I
p
(max), Timeout and Cutoff voltage. Users can type the values with
the Entry keys.