203
Operator:
1 Touch Patient Test.
Perform the patient test as usual. Once the
test is completed, the observer has to com-
plete the next steps.
2 Hand the meter back to the observer.
Observer:
3 Touch to log in again.
4 After you enter your password, touch
to proceed with the assessment.
5 Assess the operator’s performance by
touching Pass or Fail.
6 Assess the test result by touching Accept
or Reject.
7 Touch to add a comment.
8 Touch to return to the Pat. Test - OTS
screen.
The OTS information is saved together with
the test result.
Pat. Test - OTS
Observer Login
Patient Test
OTS enabled
04/19/2016
Observer Login
I-010
Observer Alan Smithee, please log
in to proceed. Otherwise the OTS
test will not be complete.
Observer
OTS passed:
Pass Fail
04/19/2016
Measurement Result OK?
Reject Accept
Comment:
Remove Strip
Remove Strip
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)