SEVEN PLUS User’s Guide | 41
HELPFUL HINTS:
• If you use the Auto Upload feature, you must register your OneTouch meter to your Receiver the
rst time you use them together or after you change the time on your Receiver. You do not need to
register each time you calibrate.
• Blood glucose information stored in the OneTouch Ultra Meter before registering it to your Receiver
will not be used for calibration. Always register a new meter before uploading ngersticks for
calibration.
• The Receiver and OneTouch Ultra meter date and time must be the same. If they are not, the
ngersticks will not be accepted by the System for calibration. Anytime you change the date/time on
your Receiver, connect the meter to the Receiver using the Connection Cable (without taking new
ngerstick readings) to match the date/time on the devices. You can upload ngerstick readings as
usual after that.