SEVEN PLUS User’s Guide | 135
APPENDIX I, OTHER ALERTS
This table shows the Other Alerts. You can also learn how to clear the Alert and when you will
be re-alerted by the Receiver for the event:
Screen Alert Title Description Re-Alert Clearing
Alerts
Re-Alert
After
Clear
Low
Battery Alert
The Receiver’s
battery is low.
Charge your Re-
ceiver as soon as
possible when
you see the Low
Battery Alert.
2 times in
the next 10
minutes
Press C but-
ton or charge
the Receiver
Every 60
minutes
“Enter BG”
Single Blood
Drop Alert
The Receiver
needs a
calibration
ngerstick. Take
a ngerstick and
enter the reading
into the Receiver.
None Press C
button
Every 15
minutes
“Enter 2 More
Startup BGs”
Repeat Double
Blood
Drop Alert
You must enter
two additional
ngerstick
readings for
Start-Up
calibration.
Every 5
minutes
Press C
button
Every 15
minutes
“Enter 1 More
Startup BG”
Shaded-Out
Blood Drop
Alert
You must enter
one additional
ngerstick
reading for Start
Up
Calibration.
Every 5
minutes
Press C
button
Every 15
minutes
11