H50/H51 Hematology Analyzer User Manual Service
- 56 -
Replace or maintain
counting baths
Drain Baths
This operation is only for technical persons
not users.
Replace optical bench
system.
Drain Flow Cell
This operation is only for technical persons
not users.
Replace
sampling
needle
and/or rins
ing head O
-
ring
Needle Dismantling
This operation allows moving the sampling
module to
a position so that it
s very
convenient
for users to
access to the
sampling
needle
Please refer to
Section 13.6.1.3
for detailed
instructions of performing needle
dismantling.
Transport analyzer to a
new place.
Park Motors
Transport analyzer to a
new place or
stop
to use
the analyzer for a long
time.
This operation will use bleach and distilled
water to do a complete cleaning of the
analyzer and then dry the whole fluidic
system.
Grease
syringe
s piston
Syringe Greasing
This operation allows moving the pistons
down in order to perform the greasing.
Please refer to
Section 13.6.1.2
for detailed
instructions of performing syringe greasing.
M
otors problems
Check Rocker
This operation controls the functionality of
the sampling module transversal axis
and
checks the motor and sensors.
Check Syringe
This option allows controlling the syringe
module functionality
and
checks the syringe
motor and sensor.
Check Needle
This option allows controlling the
functionality of the Y axis of the sampling
module, this function checks the motor and
sensors.
C
heck Valves and
Sensors
Check Sensors/Valves
Please refer to
Section 13.6.1.2
for detailed
information.
13.6.1.1 Instructions for Performing Bleach Cleaning
1) Press Bleach Cleaning, the following screen pops out.