H50/H51 Hematology Analyzer User Manual Service
- 57 -
2) Press the following screen pops out.
3) When the following screen pops out, open the fluidic door with screw driver and add 4mL of
bleach (Sodium Hypo chloride solution) into RBC bath and WBC bath respectively.
NOTE͵
The system will beep when it requires the user to pour the bleach into the baths
4) Close fluidic door and Press on the pop-out. The system will automatically perform
bleach cleaning.
13.6.1.2 Instructions for Performing Syringe Greasing
Wear rubber glove on one
hand and place a bit of
silicon grease at the tip of
the index.
Spread a thin film of
grease on each piston.
With a tool key type T20,
turn the two waste pistons in
order to spread the grease all
around the pistons.