Patient Monitor User Manual Other Functions
- 131 -
Maintain > Other Setups, and set Data Store to On or Off.
The monitor will stop storing data in the storage device under the following circumstances:
No storage devices are selected.
There is no enough space in the storage device for storing data.
The removable device is read-only.
The data storing function is deactivated.
The monitor is switched off.
The power supply is off.
22.4.3 Selecting a Storage Device
To configure the storage device, select Menu > Common Function > Data Store > Storage
Medium, and choose the storage medium from the pop-up list as desired. Internal Storage
Device and Removable Device can be selected.
When you choose Internal Storage Device as the storage medium, if configured, the storage
device name will automatically become Internal Storage Device. You may plug several
removable devices into the monitor at the same time, but only one is operative. You can select a
removable device as a working one among the plugging devices by selecting Menu > Common
Function > Data Store > Storage Device and choosing the device name from the list. By default,
the first plugged removable device is the working one.
After you configure the appropriate storage device, click exit. If the storage device is successfully
starting data storing, the monitor will be indicated by the symbol . If there is no enough
space in storage device, or the storage device is read-only/damaged, the symbol will be
displayed.
CAUTION
1 Not all the removable devices are compatible with the monitor, Use the removable
devices recommended by EDAN.
2 Do not set the read-only switch on the removable device to on when the removable
device is inserted in the monitor.
22.4.4 Reviewing Data Stored in the Storage Device
To review data stored in the storage device, select Menu > Review > History Patient. You can
choose to review the storage device as desired from the pop-up list. Choose a patient from the list
to review the data including patient information, trend graph, trend table, NIBP measurements,
arrhythmia event, alarm event, 12-lead analysis and waveform.
22.4.5 Deleting Data Stored in the Storage Device
To delete data of one patient, choose the patient from the list after selecting Menu > Review >