Patient Monitor User Manual Product Specification
- 161 -
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
Complied with IEC 60601-2-27: 2011, Sect. 201.7.9.2.9.101 b)
4), the HR value after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy: 80 bpm±1 bpm
Slow alternating ventricular bigeminy: 60 bpm±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm±1 bpm
Bidirectional systoles: 91 bpm±1 bpm
Time to Alarm for Heart
Rate alarm conditions
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
12-Lead ECG
Synchronization Analysis
Average parameters of heart beat
Time limit of P wave (ms)
A.5 RESP
Impedance between RA-LL, RA-LA
Options are lead I and II. The default is Lead II.
200 Ω to 2500 Ω (with ECG cables of 1 KΩ resistance)