Patient Monitor User Manual Default Settings
- 183 -
Alarm Level
Medium
Unit
mmHg
Filter
12.5 Hz
SYS, DIA, MAP
SYS, DIA, MAP
SYS, DIA, MAP
Alarm High Limit (ART,
P1, P2)
160, 90, 110
120, 70, 90
90, 60, 70
Alarm Low Limit (ART,
P1, P2)
90, 50, 70
70, 40, 50
55, 20, 35
Alarm High Limit (PA)
35, 16, 20
60, 4, 26
60, 4, 26
Alarm Low Limit (PA)
10, 0, 0
24, -4, 12
24, -4, 12
MAP
MAP
MAP
Alarm High Limit (CVP,
RAP, LAP, ICP)
10
4
4
Alarm Low Limit (CVP,
RAP, LAP, ICP)
0
0
0
C.10 CO
2
Default Settings
CO
2
Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Work Mode
Standby
Unit
mmHg
Apnea Alarm Time
20 s
O
2
Compensate
16%
N
2
O Compensate
0%
Water vapor Compensate
Off
Flow rate
100%
Anes Agent
0.0%
Alarm High Limit (EtCO
2
)
50
50
45
Alarm Low Limit (EtCO
2
)
25
25
30
Alarm High Limit (FiCO
2
)
4
4
4
Alarm High Limit (AwRR)
30
30
100

Table of Contents