EasyManua.ls Logo

EDAN X8 - NIBP Default Settings; TEMP Default Settings; IBP Default Settings

EDAN X8
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Default Settings
- 182 -
C.7 NIBP Default Settings
NIBP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit (SYS)
160
120
90
Alarm Low Limit (SYS)
90
70
40
Alarm High Limit (Map)
110
90
70
Alarm Low Limit (Map)
60
50
30
Alarm High Limit (DIA)
90
70
60
Alarm Low Limit (DIA)
50
40
20
Inflation value
160
140
100
Unit
mmHg
Interval
Manual
C.8 TEMP Default Settings
TEMP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit (T1)
39.0
39.0
39.0
Alarm Low Limit (T1)
36.0
36.0
36.0
Alarm High Limit (T2)
39.0
39.0
39.0
Alarm Low Limit (T2)
36.0
36.0
36.0
Alarm High Limit (TD)
2.0
2.0
2.0
Unit
C
C.9 IBP Default Settings
IBP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off

Table of Contents