Patient Monitor User Manual Alarm Information
- 45 -
Less than 10 M space is left
in the removable device.
Delete some data in
the removable device
or use another
removable device.
The removable device is
read-only.
Repair the removable
device or replace it
with a new one.
In distributed alarm system,
the monitor’s network is
disconnected.
1) Check if the
network cable is well
connected.
2) Check if the CMS
is turned on.
3) Check if the IP of
bedside monitor and
CMS are on the same
network segment.
Audio circuit connection is
abnormal, or loudspeaker
falls off.
Stop using the
monitor and notify
the manufacturer’s
service staff.
5.3 Prompts
The QRS template building required for Arr. Analysis is in
process.
The ST algorithm cannot produce valid ST value, which
may be caused by the large change in the measured value of
connected cardiogram ST or ventricular pacing.
The arrhythmia algorithm cannot analyze ECG data reliably.
V-Fib/V-Tach alarm is set to Off.
Extreme Tachycardia alarm is set to Off.
Extreme Bradycardia alarm is set to Off.
V-Tach alarm is set to Off.
Vent Brady alarm is set to Off.