Patient Monitor User Manual Alarm Information
- 45 -
Message
Cause
Alarm Level
Action Taken
Removable device is
full
Less than 10 M space is left
in the removable device.
Low
Delete some data in
the removable device
or use another
removable device.
Removable dev
read-only
The removable device is
read-only.
Low
Repair the removable
device or replace it
with a new one.
Network Disconnect
In distributed alarm system,
the monitors network is
disconnected.
Low
1) Check if the
network cable is well
connected.
2) Check if the CMS
is turned on.
3) Check if the IP of
bedside monitor and
CMS are on the same
network segment.
Audio Failed
Audio circuit connection is
abnormal, or loudspeaker
falls off.
High
Stop using the
monitor and notify
the manufacturer’s
service staff.
5.3 Prompts
Message
Cause
ECG ARR Learning
The QRS template building required for Arr. Analysis is in
process.
Unable to analyze ST
The ST algorithm cannot produce valid ST value, which
may be caused by the large change in the measured value of
connected cardiogram ST or ventricular pacing.
Unable to analyze ECG
The arrhythmia algorithm cannot analyze ECG data reliably.
V-Fib/V-Tach Off
V-Fib/V-Tach alarm is set to Off.
ExtremeTachy Off
Extreme Tachycardia alarm is set to Off.
ExtremeBrady Off
Extreme Bradycardia alarm is set to Off.
V-Tach Off
V-Tach alarm is set to Off.
Vent Brady Off
Vent Brady alarm is set to Off.