Patient Monitor User Manual Product Specification
- 163 -
Neonatal Mode
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg
Alarm Type
SYS, DIA, MAP, PR(NIBP)
Cuff Pressure Measuring
Range
0 mmHg to 300 mmHg
Pressure Resolution
1 mmHg
Maximum Mean Error
±5 mmHg
Maximum Standard Deviation
8 mmHg
Maximum Measuring Period
Adult/Pediatric
120 s
Neonate
90 s
Typical Measuring Period
20 s to 35 s (depend on HR/motion disturbance)
Dual Independent Channel Overpressure Protection
Adult
(297±3) mmHg
Pediatric
(245±3 ) mmHg
Neonatal
(147±3) mmHg
Pre-inflation Pressure
Adult Mode
Range: 80/100/120/140/150/160/180/200/220/240 mmHg
Pediatric Mode
Range: 80/100/120/140/150/160/180/200 mmHg
Neonatal Mode
Range: 60/70/80/100/120 mmHg
A.7 SpO