Patient Monitor User Manual EMC Information
- 172 -
B EMC Information
- Guidance and Manufacture’s Declaration
B.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
X8 X10 and X12 are intended for use in the electromagnetic environment specified below.
The customer or the user of X8/X10/X12 should assure that they are used in such an
environment.
Electromagnetic environment – guidance
X8 X10 and X12 use RF energy only for their
internal function. Therefore, their RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
X8 X10 and X12 are suitable for use in all
establishments, other than domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
NOTE:
The EMISSIONS characteristics of X8 X10 and X12 make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) X8 X10 and X12 might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
B.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
X8 X10 and X12 are intended for use in the electromagnetic environment specified below. The
customer or the user of X8 X10 and X12 should assure that they are used in such an
environment.