Patient Monitor User Manual Default Settings
- 178 -
C Default Settings
This appendix documents the most important default settings of your monitor as it is delivered
from the factory.
Note: If your monitor has been preconfigured according to your requirements, the settings at
delivery will be different from the default settings listed here.
C.1 Patient Information Default Settings
Patient Information Settings
Patient Type
Adult
Pace
Off
C.2 Alarm Default Settings
Alarm Settings
Pause Time
120 s
Sensor Off Alarm
Off
Alarm Latch
Off
C.3 ECG Default Settings
ECG Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit
120
160
200
Alarm Low Limit
50
75
100
ARR Analysis ThresholdValue
Low Voltage(Limb)
0.5 mV
Pause
3 s
Sustain VT
30 s
PAC Bigeminy
8/min
Pauses/min High
8/min
PVCs High
10/min
PAC Trigeminy
16/min