Patient Monitor User Manual Product Specification
- 156 -
NIBP Measurement Review
1200 sets
Arrhythmia events
Up to 200 sets
12-lead analysis Review
Up to 50 sets
Refer to Chapter 18 Review for more information about data review.
Data Storage
A single piece of patient data maximally contains the following information:
Trend graph and trend table
240 hours, resolution: 1 min
NIBP measurement review
1200 sets
Alarm review
200 sets
Arrhythmia event
200 sets
12-lead analysis review
50 sets
Full disclosure Waveforms
3- electrode/5- electrode/6- electrode:48 hours
10- electrode: 35 hours
The following storage capacity for 1G extended space is for reference:
Continuous parameter data
720 hours, resolution: 1 min
NIBP data
At least 68000 sets
Physiological alarm event
At least 4500 sets
Arrhythmia event
At least 4500 sets
Full disclosure waveforms
720 hours
Refer to Section 22.4 Storing Data in the Storage Device for more information about storing data
in the storage medium.
A.3 Wi-Fi (Optional)
IEEE
802.11b/g/n
Frequency Band
2.4 GHz ISM band
Modulation
OFDM with BPSK, QPSK, 16-QAM, and 64-QAM
802.11b with CCK and DSSS
Typical Transmit Power (±2 dBm)
17 dBm for 802.11b DSSS
17 dBm for 802.11b CCK
15 dBm for 802.11g/n OFDM

Table of Contents