Patient Monitor User Manual Product Specification
- 169 -
A.13 Interfaces
A.13.1 Analog Output (Optional)
Bandwidth (-3 dB; reference
frequency: 10 Hz)
Diagnosis/Monitor: 0.5 Hz to 40 Hz
Diagnosis 1: 0.05 Hz to 40 Hz
Surgery: 1 Hz to 20 Hz
Enhanced: 2 Hz ~18 Hz
Customized: When Low-pass Filter < 40 Hz, Bandwidth
is High-pass Filter ~ Low-pass Filter; When Low-pass
Filter > 40 Hz, Bandwidth is High-pass ~40 Hz.
Maximum Transmission Delay
(Diagnosis Mode)
PACE Rejection/ Enhancement
Consistent with the calculation leads.
Compliant with Standard and
Directive
Complies with the requirements in terms of short circuit
protection and leakage current in EN60601-1.
NOTE:
While using analog output, set the calculation lead as following:
1) In 3-Electrode mode, set to Lead I, Lead II, or Lead III.
2) In 5-Electrode mode, set to Lead I, Lead II, Lead III or Lead V.
3) In 6-Electrode mode, set to I, II, III, and leads corresponding to Va, Vb.
4) In 10-Electrode mode, set to Lead I, Lead II, Lead III or Lead V1~V6.
A.13.2 Defibrillator Synchronization (Optional)
35 ms (R-wave peak to leading edge of pulse)
High level: 3.5 V to 5.5 V, providing a maximum of 1 mA
output current;
Low level: < 0.5 V, receiving a maximum of 5 mA input
current