Patient Monitor User Manual Product Specification
- 166 -
A.11 CO
2
Complies with ISO 80601-2-55: 2011.
Intended
Patient
Adult, pediatric, neonatal
Measure
Parameters
EtCO
2
, FiCO
2
, AwRR
Unit
mmHg, %, kPa
Measuring
Range
CO
2
0 mmHg to 150 mmHg (0 % to 20%)
AwRR
2 rpm to 150 rpm
Resolution
EtCO
2
1 mmHg
FiCO
2
1 mmHg
AwRR
1 rpm
Accuracy
EtCO
2
2 mmHg, 0
mmHg to 40
mmHg
Respiratory
rate 60
rpm
Typical conditions:
Ambient temperature: (25 3)
C
Barometric pressure: (760 10)
mmHg
Balance gas: N
2
Sample gas flowrate: 100 ml/min
5% of reading, 41
mmHg to 70
mmHg
8% of reading, 71
mmHg to 100
mmHg
10% of reading,
101 mmHg to 150
mmHg
± 12% of reading
or ± 4 mmHg,
whichever is
greater
Respiratory
rate > 60
rpm
All conditions
AwRR
1 rpm
Drift of
Measure
Accuracy
Meets the requirements of the measure accuracy
Sample Gas
Flowrate
50 ml/min, 70 ml/min or 100 ml/min(default), accuracy: 15 ml/min
Warm-upTime
Display reading within 20 s; reach to the designed accuracy within 2 minutes.
Rise Time
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Response
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)