EasyManua.ls Logo

EDAN X8 - Page 189

EDAN X8
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual EMC Information
- 176 -
Table 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test
frequenc
y (MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
Immuni
ty test
level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704-787
LTE Band
13, 17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400-2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
5240
5100-5800
WLAN
802.11 a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The
1 m test distance is permitted by IEC 61000-4-3.

Table of Contents

Related product manuals