Patient Monitor User Manual Alarm Information
- 43 -
Message
Cause
Alarm Level
Action Taken
IBP Sensor Err
Malfunction in the IBP
sensor or in the extension
cable.
Medium
Replace the IBP
sensor or the
extension cable.
YY Comm Fail (YY
stands for the IBP
label name: Art, PA,
CVP, RAP, LAP,
ICP, P1 and P2)
IBP module failure or
communication failure
High
Stop measuring
function of IBP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
C.O.
C.O. TI Sensor Off
C.O. TI sensor not connected
Low
Insert injective
temperature sensor.
C.O. TB Sensor Off
C.O. TB sensor not
connected
Low
Insert TB sensor.
C.O. Comm Fail
C.O. module failure or
communication failure
High
Stop measuring of
C.O. module, or
notify biomedical
engineer or
Manufacturer’s
service staff.
C.O. TEMP Out Of
Range
TI/TB measuring value is
beyond measuring range.
High
Please check TI/TB
sensor.
CO
2
CO
2
Comm Fail
CO
2
module failure or
communication failure
High
Check if the water
tray has been fixed.
CO
2
Check Adapter
The water trap is
disconnected or not properly
connected.
Low
Properly connect the
water trap.
CO
2
Sensor Over
Temp
CO
2
sensor temperature
exceeds +40 C.
High
Stop using
measuring function
of CO
2
module,
notify biomedical
engineer.
CO
2
Sensor Faulty
CO
2
module failure
High
FiCO
2
Overrange
The FiCO
2
concentration
exceeds the measurement
range.
High
Please check the
monitor or patient
status and adjust the