STP-iS1607 Series Turbomolecular Pump
MT-94E-001-C
Page 119
5
SERIAL COMMUNICATION PROTOCOL
Parameter
Item
Data format
Remarks
11
Pump Overload Warning
Rotational speed setting
16-bits hexadecimal
coded ASCII
Default 0000
hex
(0.0%)
12
Serial communication
time out setting (Unit: sec.)
16-bits hexadecimal
coded ASCII
Default 003C
hex
(60 sec.)
13
[System reservation]
88-bits hexadecimal
coded ASCII
*2
*1
Valid only with TMS specification
*2
Assign the parameter value the reading data of ReadOptionFunc (?=) or F
hex
.
Refer to Table 29, "Parameter setting value" for each parameter setting value.
The value is not reflected when a parameter is out of a setting range.
Example:
1. Input operation port: 01
hex
= I/O REMOTE
2. TMS Option: FF
hex
= DISABLE
4. Second Damage Limit Option: 00
hex
= ENABLE
5. First Damage Limit Warning: 00
hex
= ENABLE
6. Pump Runtime Over Warning: FF
hex
= DISABLE
7. Pump Runtime Over Warning hours: 000003E8
hex
= 1,000 (×100 hours)
8. Imbalance Warning: 00
hex
= ENABLE
9. Pump Overload Warning: FF
hex
= DISABLE
10. Pump Overload Warning Motor current: 03E8
hex
= 1,000 (×0.1%)
11. Pump Overload Warning Rotational speed: 0000
hex
= 0 (×0.1%)
12. Serial communication time out: 003C
hex
= 60 sec.
Parameter
1
2
3
4
5
ASCII
0
1
F
F
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
0
0
0
0
HEX
30
31
46
46
30
30
30
30
Parameter
6
7
8
ASCII
F
F
0
0
0
0
0
3
E
8
0
0
HEX
46
46
30
30
30
30
30
33
45
38
30
30
Parameter
9
10
11
12
13
ASCII
F
F
0
3
E
8
0
0
0
0
0
0
3
C
*1
*1
*1
*1
*1
*1
HEX
46
46
30
33
45
38
30
30
30
30
30
30
33
43

Table of Contents

Related product manuals