STP-iS1607 Series Turbomolecular Pump
MT-94E-001-C
Page 147
7
SAFETY FUNCTIONS
Referred
Section
5.2.5
*1
The driver output stops and the rotor continue rotating by inertia.
It may take several hours to stop.
To stop the STP pump quickly, close the vacuum valve at the outlet port
flange and introduce gas from the purge port into the STP pump.
Table
42 - Error List
(continued)
Countermeasures
Contact Service office.
1. Send
a communication
command periodically at
intervals of less than
setting value
.
2. Check the communication
cable.
3. Contact Service office.
Contact Service office.
Overhaul is needed
.
Contact Service office.
Probable causes
Failure of the
control unit.
1. Insufficient user
application setting
.
2. Disconnection
or break
of
the communication cable.
3. Failure of the
control unit.
Failure of the
control unit.
Frequency of "Disturbance"
or
"Power
Failure" error
.
Occurrence
condition
Abnormal power
circuit
.
Communication
failure of the input
operation port
during
acceleration
and
normal operation
Abnormal rotatinal
speed detection
Damage point
excees setting
value
Pump operation
Free run
*1
Decelerat
e and
stop
Free run
*1
Decelerat
e and
stop
(Error
message
)
Power Supply
Fail
Serial Com. Fail
Speed Pulse
Lost
START NOT
ALLOWED
"FAILURE"
LED
Steady
red

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards STP-iS1607 Series and is the answer not in the manual?

Edwards STP-iS1607 Series Specifications

General IconGeneral
BrandEdwards
ModelSTP-iS1607 Series
CategoryWater Pump
LanguageEnglish

Related product manuals