STP-iS1607 Series Turbomolecular Pump
*1
The magnetic bearing output stops and the rotor is descended on the touch down bearing.
*2
The driver output stops and the rotor continue rotating by inertia.
It may take several hours to stop.
To stop the STP pump quickly, close the vacuum
valve at the outlet port flange and introduce gas from the purge port into the STP pump.
Table
41 - Error List
(continued)
1. Check whether or
not a
power failure has occurred.
2. Connect the power cable
correctly.
3. Replace the power cable.
4. Set the
rated
voltage ±
10 %.
1. Check the volume of water
and water temperature with
water cooling unit.
In order
to recover, turn off the
power (switch "OFF"), and
then switch "ON" after
sufficient cooling of the
control
unit.
(isolate input power supply
3 minutes or more
)
2. Contact Service office.
1. Failure of the STP
pump.
2. Failure of the
control unit.
Disconnection of the
motor
temperature
sensor
.
Failure of the
control unit.
2. Incorrect connection of the
power cable.
3. Failure of
the power cable.
4. Failure of the power
supply vlotage.
1. Insufficient cooling of the
control unit
.
2. Failure of
the control unit.
Excessive
magnetic bearing
electric current
continues for 30
sec.
Abnormal
motor
temperature
detection
Rotaional speed
exceeds rated
speed
Insufficient power
supply
Overheat inside
the control unit
.