Fault removal multi N/C 2100S
94
13 No connection to the autosampler
Cause Remedy
sample not switched on
connection cable not connected or faulty
switch on sampler and initialize analyzer
check connection cable
15 no O
2
pressure at HT furnace
Cause Remedy
O
2
no connection present or faulty Connect the carrier gas to HT furnace (check 4 6
bar preliminary pressure)
16 TC lock failure (folding head)
Cause Remedy
The automatic lock does not open Check the gas pressure of the auxiliary gas for the
locks (4 to 6 bar preliminary pressure required)
Check the hose connections of the lock
20
21
22
26
no connection to optics (NDIR)
CRC error optics
status error optics
optics error; incorrect command return
Cause Remedy
Communication errors initialize analyzer
NDIR detector faulty inform Service
24 Optics error, analog values outside of range
Cause Remedy
the analog values of the detector are outside
the working range
check the quality of the carrier gas
initialize analyzer and check analog values via
component test
27 Optics error, analog values outside of range
Cause Remedy
the analog values of the detector are outside
the working range
check quality of the carrier gas
for solids methods and connection of HT 1300
carrier gas flow > intake flow
initialize analyzer and check analog values via
component test
30 No connection to the N sensor
Cause Remedy
CLD is not switched on
connection cable not connected or faulty
incorrect connection
switch on CLD
check connection cable
check connection

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Analytik Jena multi N/C 2100S and is the answer not in the manual?

Endress+Hauser Analytik Jena multi N/C 2100S Specifications

General IconGeneral
BrandEndress+Hauser
ModelAnalytik Jena multi N/C 2100S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals